CY30 pocket oxygen detection alarm operating procedures

News Related Keywords: No tags.

CY30 pocket oxygen detection alarm

CY30 Pocket Oxygen Detection for coal mine tunnel continuous monitoring of oxygen in the environment, etc. O2 oxygen concentration detection equipment. When the oxygen concentration is too low, sound, light, vibration, and light bar alarms are automatically emitted. This product is an intrinsically safe and explosion-proof type, explosion-proof mark is ExibdI.

The CY30 Pocket Oxygen Detector uses an electrochemical sensor whose structure consists of three electrodes in the electrolyte.

CY30 Pocket Oxygen Detection Alarm Application Location: Coal mine underground personnel carrying, industrial and civil environment Oxygen measurement cable type with probe extension line, suitable for the hypoxia measurement of tanks, pipelines and underground tunnels and other places.

CY30 pocket oxygen detection alarm operating procedures

First, the scope

This procedure is used for the operation and maintenance of the CY30 Pocket Oxygen Detection Alarm, aiming to regulate the operation and maintenance of the personnel.

Second, terms and definitions

Combustible gas alarm : It is mainly used for detecting combustible gas in the air. It can be divided into fixed combustible gas alarm and portable combustible gas alarm.

Third, responsibility

1 The post operator is responsible for the operation and maintenance of the fixed combustible gas alarm.

2 Gas transmission stations and patrol patrol squadron operators responsible for the operation and maintenance of portable combustible gas alarms;

3 The safety officer in the management area is responsible for the detection of combustible gas alarms;

4 The security officer in the management area is responsible for the archival management of the data;

5 The competent department is responsible for external maintenance contact, project supervision, and organization acceptance of external maintenance.

Fourth, the operating procedures

1 gas detector operating procedures

Power on operation: When the instrument is powered on ( 12-28VDC ), it will start working. First enter the self-test state. The LED will display the version number of the instrument software and sensor software. Then the instrument enters the warm-up state and the leftmost decimal point flashes, three minutes warm up After that, the instrument enters normal operation mode.

2 zero operation

( 1 ) Zeroing operation with magnetic bar: In normal operation mode, the instrument enters the programming mode after the CH1 point is set by a magnetic bar. After entering this mode for 5 seconds, the first function option is “zeroing- 0. ” Display “ 0. ” in the status bit and start counting down for 10 seconds. Before the countdown is completed, use the magnetic rod to point CH2 and the instrument will start zero calibration. After the zero calibration is over, the instrument will automatically return to the normal display mode.

( 2 ) Use key to perform zero adjustment operation: First, open the shell of the instrument, press MODE key, and then press “ ENTER ” key after “ MODE ”, “ + ”, “ - ”, “ + ” key to confirm, and the instrument enters programming mode. , and then press " + " or " - " key to switch function options to "zero -0. " At this time, the status bit shows " 0 " and starts counting down for 10 seconds. Press the confirmation key before the countdown is completed to start zeroing. Return to normal display mode automatically.

3 calibration operation

( 1 ) Use magnetic bar to perform calibration operation: In normal operation mode, the instrument enters programming mode after the CH1 point is set by a magnetic bar , and then the magnetic bar point CH1 is used to switch to the function option “Calibration- C ”.

At this time, the status bit shows “ C. ” and starts counting down for 10 seconds. Before the countdown is completed, use the magnetic bar to point CH2 and the instrument starts to calibrate. After starting the calibration procedure, the instrument will automatically zero and display the flashing “ 0. ” in the status bit . ”, and start timing. After the zero adjustment, the instrument enters the calibration state. The upper line of the LED shows “ C.XXX ”. At this time, it is necessary to provide the instrument with the standard gas related to the calibration gas concentration. During the calibration process, the calibration value will change from the digital zero until the calibration automatically ends. The instrument will automatically return to the normal display mode after the calibration is completed;

( 2 ) Use key to perform calibration operation: firstly open the shell of the instrument, press MODE key , then press “ MODE ” “ + ”, “ - ”, “ + ” key, press ENTER key to confirm, the instrument enters programming mode, and then Press " + " or " - " key to switch function options to "Calibration- C ", then the status bit displays " C. " and starts counting down for 10 seconds. Before the countdown is completed, press the confirmation key to start calibration and start the calibration process. The instrument is automatically zeroed first , and the flashing “ 0. ” is displayed in the status bit. After the zero adjustment is completed, the instrument enters the calibration state. The LED on the top line shows “ C.XXX ”. At this time, the instrument needs to be supplied with the calibration gas. Concentration-related standard gas. During the calibration process, the calibration value will change from the digital zero until the calibration automatically ends. After calibration, the instrument automatically returns to normal display mode.

Fifth, portable combustible gas alarm operating procedures

1 In non-hazardous locations (where the air is clean), press the key to open the alarm;

2 The warm-up and "zero" adjustments are completed in about 1 minute. Check the battery power while warming up;

3 After the warm-up is completed, it can enter the dangerous place for testing;

4 When the test is over, keep pressing the key until the countdown is completed and the power is turned off.

This article is edited and compiled by China Rescue Equipment Network. Please indicate the link.