High and low temperature alternating damp heat test box fault display code and processing method

Preface: In the long-term use of high-low temperature alternating test chamber, due to various reasons, it may be faulty. The relevant fault will appear related to the code on the touch screen controller, and the customer must fault according to the displayed fault code. Analyze and eliminate.

Please check the fault message on the controller for the fault status. The troubleshooting instructions are as follows: If the user knows some electrician skills, he can deal with the supplier technician and deal with it himself. If the user does not have a certain electrician foundation, please notify the first time. The factory arranges technicians to overhaul and promptly repair the high and low temperature test chambers so as to avoid causing more serious failures and make repair and repair of high and low temperature test chambers more costly.

ERROR1

CM1 compressor overcurrent protection jump

the reason

1

Phase loss, three-phase 380V voltage breaks one or two phases

2

High voltage

3

Compressor card cylinder

4

Condenser exhaust fan does not turn

Exclusion

1

Repair the three-phase circuit and check the power supply of the equipment

2

Supply 380V or increase the voltage regulator

3

Replace compressor

4

Replace condenser exhaust fan

ERROR2

CM1 compressor overpressure protection trip

the reason

1

Excessive ambient temperature

2

Condenser exhaust fan does not work, damaged

3

Refrigerant solenoid valve does not open

4

Too much refrigerant

Exclusion

1

Change the ambient temperature and keep it below 30 degrees Celsius

2

Replace condenser fan

3

Check if the solenoid valve is clogged, replace the solenoid valve

4

Drain some refrigerant (refrigerant)

ERROR3

CM2 compressor overcurrent protection jump

the reason

1

Phase loss, three-phase 380V voltage breaks one or two phases

2

High voltage

3

Compressor card cylinder

4

Condenser exhaust fan does not turn

5

The first stage compressor has too little refrigerant

Exclusion

1

Repair the three-phase circuit and check the power supply of the equipment

2

Supply 380V or increase the voltage regulator

3

Replace compressor

4

Replace condenser exhaust fan

5

Fill compressor refrigerant

ERROR4

CM2 compressor overpressure protection trip

the reason

1

Excessive ambient temperature

2

Condenser exhaust fan does not work, damaged

3

Refrigerant solenoid valve does not open

4

Too much refrigerant

Exclusion

1

Change the ambient temperature and keep it below 30 degrees Celsius

2

Replace condenser fan

3

Check if the solenoid valve is clogged, replace the solenoid valve

4

Drain some refrigerant (refrigerant)

ERROR5

Abnormal fan in the box

the reason

1

Phase loss, three-phase 380V voltage breaks one or two phases

2

High voltage

3

Damaged motor bearings

4

The motor coil has burned out

5

Thermal relay burnout

Exclusion

1

Repair three-phase circuit

2

Supply 380V or increase voltage regulator

3

Replace motor bearing

4

Repair motor coil or replace motor

5

Replacing the thermal relay

ERROR6

Cooling tower anomaly

the reason

1

High voltage

2

Damaged motor bearings

3

The motor coil has burned out

4

Thermal relay burnout

Exclusion

1

Supply AC220V voltage or increase voltage regulator

2

Replace motor bearing

3

Repair motor coil or replace motor

4

Replacing the thermal relay

ERROR7

Temperature protection in the box

the reason

1

Test box temperature exceeds the set protection temperature range

2

Overtemperature protector is damaged

3

Broken temperature sensor circuit

4

Solid state relay damage

Exclusion

1

Check the protection temperature is correct, reset the protection temperature

2

Replace over-temperature protector

3

Replace new temperature sensor to complete overhaul

4

Replace solid state relay

ERROR8

Humidification tube air protection

the reason

1

Humidity bucket no water or too low

2

Anti-dry protection protector damaged

3

Solid state relay damage

Exclusion

1

Check whether the pump is working (pumping), whether the water pipe is blocked, the water level of the storage tank is reasonable, and adjust the water level box upward so that the water completely submerges the humidifying pipe.

2

Replace anti-dry protection device

3

Replace solid state relay

ERROR9

Water tank water shortage

the reason

1

Water level in storage tank is too low

2

Water level floater bite (caught)

3

Water float damage

Exclusion

1

Supplement pure water, check whether the automatic water supply device is normal

2

Check the cleanliness of the tank and clean the storage tank

3

Replace water level float

ERROR10

Power supply phase or reverse phase

the reason

1

Total power supply phase loss (one or two phases of three phase 380V voltage)

2

Three-phase power supply (R S T) reversed

3

Phase sequence protector is damaged

Exclusion

1

Repair three-phase circuit

2

Replace the two middle fire wires of the three-phase power line with each other

3

Replace phase sequence protector